Noninterchangeability of Vessel Density Measurements Across Next-Generation SS-OCTA Systems.


ABSTRACT:

Purpose

Comparing swept-source optical coherence tomography angiography (OCTA) vessel density in the superficial and deep capillary plexuses across leading OCTA systems.

Methods

OCTAs were acquired from healthy participants using two OCTA systems (BMizar 400 kHz and Yalkaid 100 kHz, TowardPi, Beijing, China) with scan sizes of 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm2. The vessel density was quantified using the manufacturer's software.

Results

Across 30 subjects comprising 60 eyes and a total of 225 OCTA scans, significant interdevice differences were observed at the Early Treatment Diabetic Retinopathy Study sector level in all eight scan size × plexus comparisons (three-way analysis of variance; P < 1.16 × 10e-4). The mean BMizar vessel density values for the superfic