Feasibility of 12-lead ECG reconstruction at rest and during adenosine stress perfusion inside a 3-Tesla MR scanner.
Ontology highlight
ABSTRACT: Background Electrocardiographic (ECG) devices in magnetic resonance (MR) scanners have a narrow bandwidth and suffer from signal distortion, preventing measurement of QRS/ QT durations, or detection of ST/ T-wave changes during ischaemia. We integrated an external device with novel post-processing software into a 3 Tesla (T) scanner to assess feasibility and accuracy for reconstructing 12-lead ECGs prior to and during adenosine infusion. Methods Reference 12-lead ECGs were first recorded outside the MR-environment. Using three MR-safe electrode patches, in-bore ECGs were then recorded prior to and during adenosine infusion. Post-processing removed artefacts and 12-lead ECGs were reconstructed from the raw signal using a subject-specific matrix. In-bore and reference ECGs were compared usin