Initial clinical experience with the HeartMate II axial-flow left ventricular assist device.


ABSTRACT: The redesigned HeartMate II, an axial-flow left ventricular assist device, is simpler, smaller, and easier to operate than are pulsatile pumps. These design characteristics should make the HeartMate II more reliable and durable and broaden the eligible population base. We implanted the HeartMate II in 43 patients (average age, 42 yr). The indication for use was bridge-to-heart transplantation in 26 patients and destination therapy in 17. The average duration of device support was 258 days (range, 1-761 days), and cumulative duration, more than 31 patient-years. Hemodynamic function improved in all patients during support. By 48 hours after implantation, the mean cardiac index had increased from 1.9+/-0.27 L/(min.m(2)) (baseline) to 3.5+/-0.8 L/(min.m(2)), and the pulmonary capillary wedge