Base-pair neutral homozygotes can be discriminated by calibrated high-resolution melting of small amplicons.
Ontology highlight
ABSTRACT: Genotyping by high-resolution melting analysis of small amplicons is homogeneous and simple. However, this approach can be limited by physical and chemical components of the system that contribute to intersample melting variation. It is challenging for this method to distinguish homozygous G::C from C::G or A::T from T::A base-pair neutral variants, which comprise approximately 16% of all human single nucleotide polymorphisms (SNPs). We used internal oligonucleotide calibrators and custom analysis software to improve small amplicon (42-86 bp) genotyping on the LightScanner. Three G/C (PAH c.1155C>G, CHK2 c.1-3850G>C and candidate gene BX647987 c.261+22,290C>G) and three T/A (CPS1 c.3405-29A>T, OTC c.299-8T>A and MSH2 c.1511-9A>T) human single nucleotide variants were analyzed. Calibration
SUBMITTER: Gundry CN
PROVIDER: S-EPMC2425497 | biostudies-literature | 2008 Jun
REPOSITORIES: biostudies-literature
ACCESS DATA