Comparative clinical study of canine and feline total blood cell count results with seven in-clinic and two commercial laboratory hematology analyzers.


ABSTRACT:

Background

A CBC is an integral part of the assessment of health and disease in companion animals. While in the past newer technologies for CBC analysis were limited to large clinical pathology laboratories, several smaller and affordable automated hematology analyzers have been developed for in-clinic use.

Objectives

The purpose of this study was to compare CBC results generated by 7 in-clinic laser- and impedance-based hematology instruments and 2 commercial laboratory analyzers.

Methods

Over a 3-month period, fresh EDTA-anticoagulated blood samples from healthy and diseased dogs (n=260) and cats (n=110) were analyzed on the LaserCyte, ForCyte, MS45, Heska CBC, Scil Vet ABC, VetScan HMT, QBC Vet Autoread, CELL-DYN 3500, and ADVIA 120 analyzers. Results were compared