Reverse current pulse method to restore uniform concentration profiles in ion-selective membranes. 1. Galvanostatic pulse methods with decreased cycle time.


ABSTRACT: The applications of ion-selective electrodes (ISEs) have been broadened through the introduction of galvanostatic current pulse methods in potentiometric analysis. An important requirement in these applications is the restoration of the uniform equilibrium concentration profiles in the ISE membrane between each measurement. The simplest restoration method is zero-current relaxation, in which the membrane relaxes under open-circuit conditions in a diffusion-controlled process. This paper presents a novel restoration method using a reverse current pulse. An analytic model for this restoration method is derived to predict the concentration profiles inside ISE membranes following galvanostatic current pulses. This model allows the calculation of the voltage transients as the membrane voltage r

SUBMITTER: Zook JM 