Patient-ventilator interaction during neurally adjusted ventilatory assist in low birth weight infants.
Ontology highlight
ABSTRACT: Neurally adjusted ventilatory assist (NAVA), a mode of mechanical ventilation controlled by diaphragmatic electrical activity (EAdi), may improve patient-ventilator interaction. We examined patient-ventilator interaction by comparing EAdi to ventilator pressure during conventional ventilation (CV) and NAVA delivered invasively and non-invasively. Seven intubated infants [birth weight 936 g (range, 676-1266 g); gestational age 26 wk (range, 25-29)] were studied before and after extubation, initially during CV and then NAVA. NAVA-intubated and NAVA-extubated demonstrated similar delays between onset of EAdi and onset of ventilator pressure of 74 +/- 17 and 72 +/- 23 ms (p = 0.698), respectively. During CV, the mean trigger delays were not different from NAVA, however 13 +/- 8.5% of ventilato
SUBMITTER: Beck J
PROVIDER: S-EPMC2762820 | biostudies-literature | 2009 Jun
REPOSITORIES: biostudies-literature
ACCESS DATA