Comprehensive assessment of myocardial perfusion defects, regional wall motion, and left ventricular function by using 64-section multidetector CT.


ABSTRACT:

Purpose

To evaluate the accuracy of 64-section multidetector computed tomography (CT) for the assessment of perfusion defects (PDs), regional wall motion (RWM), and global left ventricular (LV) function.

Materials and methods

All myocardial infarction (MI) patients signed informed consent. The IRB approved the study and it was HIPAA-compliant. Cardiac multidetector CT was performed in 102 patients (34 with recent acute MI and 68 without). Multidetector CT images were analyzed for myocardial PD, RWM abnormalities, and LV function. Global LV function and RWM were compared with transthoracic echocardiography (TTE) by using multidetector CT. PD was detected by using multidetector CT and was correlated with cardiac biomarkers and single photon emission CT (SPECT) myocardial perfu

SUBMITTER: Cury RC 

PROVIDER: S-EPMC2797649 | biostudies-literature | 2008 Aug