Relative plasma volume monitoring during hemodialysis AIDS the assessment of dry weight.
Ontology highlight
ABSTRACT: Among hemodialysis patients, the assessment of dry weight remains a matter of clinical judgment because tests to assess dry weight have not been validated. The objective of this study was to evaluate and validate relative plasma volume (RPV) monitoring as a marker of dry weight. We performed RPV monitoring using the Crit-Line monitor at baseline and at 8 weeks in 150 patients participating in the Dry-Weight Reduction in Hypertensive Hemodialysis Patients Trial. The intervention group of 100 patients had dry weight probed, whereas 50 patients served as time controls. RPV slopes were defined as flat when they were less than the median (1.33% per hour) at the baseline visit. Among predominantly (87%) black hemodialysis patients, we found that flat RPV slopes suggest a volume-overloaded state
SUBMITTER: Sinha AD
PROVIDER: S-EPMC2819307 | biostudies-literature | 2010 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA