Electrochemical quantification of serotonin in the live embryonic zebrafish intestine.


ABSTRACT: We monitored real-time in vivo levels of serotonin release in the digestive system of intact zebrafish embryos during early development (5 days postfertilization, dpf) using differential pulse voltammetry with implanted carbon fiber microelectrodes modified with carbon nanotubes dispersed in nafion. A detection limit of 1 nM, a linear range between 5 and 200 nM, and a sensitivity of 83.65 nA x microM(-1) were recorded. The microelectrodes were implanted at various locations in the intestine of zebrafish embryos. Serotonin levels of up to 29.9 (+/-1.13) nM were measured in vivo in normal physiological conditions. Measurements were performed in intact live embryos without additional perturbation beyond electrode insertion. The sensor was able to quantify pharmacological alterations in seroto