Agreement, precision, and accuracy of epifluorescence microscopy methods for enumeration of total bacterial numbers.


ABSTRACT: To assess interchangeability of estimates of bacterial abundance by different epifluorescence microscopy methods, total bacterial numbers (TBNs) determined by most widely accepted protocols were statistically compared. Bacteria in a set of distinctive samples were stained with acridine orange (AO), 4'-6-diamidino-2-phenylindole (DAPI), and BacLight and enumerated by visual counting (VC) and supervised image analysis (IA). Model II regression and Bland-Altman analysis proved general agreements between IA and VC methods, although IA counts tended to be lower than VC counts by 7% on a logarithmic scale. Distributions of cells and latex beads on polycarbonate filters were best fitted to negative binomial models rather than to Poisson or log-normal models. The fitted models revealed higher prec