Unknown

Dataset Information

Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results.


ABSTRACT:

Background

To introduce a novel method of patient positioning for high precision intracranial radiotherapy.

Methods

An infrared(IR)-array, reproducibly attached to the patient via a vacuum-mouthpiece(vMP) and connected to the table via a 6 degree-of-freedom(DoF) mechanical arm serves as positioning and fixation system. After IR-based manual prepositioning to rough treatment position and fixation of the mechanical arm, a cone-beam CT(CBCT) is performed. A robotic 6 DoF treatment couch (HexaPOD) then automatically corrects all remaining translations and rotations. This absolute position of infrared markers at the first fraction acts as reference for the following fractions where patients are manually prepositioned to within +/- 2 mm and +/- 2 degrees of this IR reference posit

SUBMITTER: Wilbert J 

PROVIDER: S-EPMC2890022 | biostudies-literature | 2010 May

REPOSITORIES: biostudies-literature