Toward an implantable functional electrical stimulation device to correct strabismus.


ABSTRACT:

Purpose

To investigate the feasibility of electrically stimulating the lateral rectus muscle to recover its physiologic abduction ability in cases of complete sixth cranial (abducens) nerve palsy.

Methods

In the feline lateral rectus muscle model, the effects of a charge-balanced, biphasic, current-controlled stimulus on the movement of the eye were investigated while stimulation frequency, amplitude, and pulse duration was varied. Eye deflection was measured with a force transducer. Denervated conditions were simulated by injection of botulinum toxin A.

Results

Three chemically denervated and 4 control lateral rectus muscles were analyzed. In control lateral rectus muscles, the minimum fusion frequency was approximately 170 Hz, and the maximum evoked abduction was 27

SUBMITTER: Velez FG 

PROVIDER: S-EPMC2908366 | biostudies-literature | 2009 Jun

REPOSITORIES: biostudies-literature