ABSTRACT: Background
Diffusive clearance depends on blood and dialysate flow rates and the overall mass transfer area coefficient (K(o)A) of the dialyzer. Although K(o)A should be constant for a given dialyzer, urea K(o)A has been reported to vary with dialysate flow rate possibly because of improvements in flow distribution. This study examined the dependence of K(o)A for urea, phosphate and β(2)-microglobulin on dialysate flow rate in dialyzers containing undulating fibers to promote flow distribution and two different fiber packing densities.Methods
Twelve stable haemodialysis patients underwent dialysis with four different dialyzers, each used with a blood flow rate of 400 mL/min and dialysate flow rates of 350, 500 and 800 mL/min. Clearances of urea, phosphate and β(2)-microglob
SUBMITTER: Bhimani JP
PROVIDER: S-EPMC2989792 | biostudies-literature | 2010 Dec
REPOSITORIES: biostudies-literature