Unknown

Dataset Information

Semi-automated non-target processing in GC × GC-MS metabolomics analysis: applicability for biomedical studies.


ABSTRACT: Due to the complexity of typical metabolomics samples and the many steps required to obtain quantitative data in GC × GC-MS consisting of deconvolution, peak picking, peak merging, and integration, the unbiased non-target quantification of GC × GC-MS data still poses a major challenge in metabolomics analysis. The feasibility of using commercially available software for non-target processing of GC × GC-MS data was assessed. For this purpose a set of mouse liver samples (24 study samples and five quality control (QC) samples prepared from the study samples) were measured with GC × GC-MS and GC-MS to study the development and progression of insulin resistance, a primary characteristic of diabetes type 2. A total of 170 and 691 peaks were quantified in, respectively, the GC-MS and GC × GC-MS

SUBMITTER: Koek MM 

PROVIDER: S-EPMC3040320 | biostudies-literature | 2011 Mar

REPOSITORIES: biostudies-literature