ABSTRACT: Purpose
Current office blood pressure measurement (OBPM) is often not executed according to guidelines and cannot prevent the white-coat effect. Serial, automated, oscillometric OBPM has the potential to overcome both these problems. We therefore developed a 30-minute OBPM method that we compared with daytime ambulatory blood pressure.Methods
Patients referred to a primary care diagnostic center for 24-hour ambulatory blood pressure monitoring (ABPM) had their blood pressure measured using the same validated ABPM device for both ABPM and 30-minute OBPMs. During 30-minute OBPM, blood pressure was measured automatically every 5 minutes with the patient sitting alone in a quiet room. The mean 30-minute OBPM (based on t = 5 to t = 30 minutes) was compared with mean daytime ABPM
SUBMITTER: van der Wel MC
PROVIDER: S-EPMC3056860 | biostudies-literature | 2011 Mar-Apr