ABSTRACT: Objective
Intracardiac beating-heart procedures require the introduction and exchange of complex instruments and devices. To prevent potential complications such as air embolism and bleeding, a universal cardioport was designed and tested.Methods
The design consists of a port body and a series of interchangeable sleeves. The port uses a fluid purging system to remove air from the instrument before insertion into the heart, and a valve system minimizes blood loss during instrument changes.Results
The cardioport was tested ex vivo and in vivo in pigs (n = 5). Beating-heart procedures, such as septal defect closure and mitral valve repair, were modeled. Ex vivo trials (n = 150) were performed, and no air emboli were introduced using the port. In comparison, air emboli