Superior fixation of pegged trabecular metal over screw-fixed pegged porous titanium fiber mesh: a randomized clinical RSA study on cementless tibial components.


ABSTRACT:

Background and purpose

Lasting stability of cementless implants depends on osseointegration into the implant surface, and long-term implant fixation can be predicted using radiostereometric analysis (RSA) with short-term follow-up. We hypothesized that there would be improved fixation of high-porosity trabecular metal (TM) tibial components compared to low-porosity titanium pegged porous fiber-metal (Ti) polyethylene metal backings.

Methods

In a prospective, parallel-group, randomized unblinded clinical trial, we compared cementless tibial components in patients aged 70 years and younger with osteoarthritis. The pre-study sample size calculation was 22 patients per group. 25 TM tibial components were fixed press-fit by 2 hexagonal pegs (TM group) and 25 Ti tibial components