Pseudo-outbreak of extremely drug-resistant pseudomonas aeruginosa urinary tract infections due to contamination of an automated urine analyzer.


ABSTRACT: By the end of May 2010, an increase in the number of urine specimens that were culture positive for extremely drug-resistant (XDR) Pseudomonas aeruginosa was observed in our 800-bed university hospital. This led to an infection control alert. No epidemiological link between the patients and no increase in the frequency of XDR P. aeruginosa in non-urine samples were observed. Therefore, a pseudo-outbreak due to analytical contamination in the laboratory was rapidly suspected. A prospective and retrospective search of cases was initiated, and the sampling of the automated urine analyzers used in the laboratory was performed. Antibiotypes were determined by disc diffusion, and genotypes were determined by pulsed-field gel electrophoresis (PFGE). From February to July 2010, 17 patients admitte

SUBMITTER: Hallin M 

PROVIDER: S-EPMC3295159 | biostudies-literature | 2012 Mar