Comparison of the Microflex LT and Vitek MS systems for routine identification of bacteria by matrix-assisted laser desorption ionization-time of flight mass spectrometry.
Ontology highlight
ABSTRACT: This study compared the performance of three matrix-assisted laser desorption ionization-time of flight mass spectrometry systems: Microflex LT (Bruker Daltonics, Bremen, Germany), Vitek MS RUO (Axima Assurance-Saramis database; bioMérieux, Marcy l'Etoile, France), and Vitek MS IVD (bioMérieux). A total of 1,129 isolates, including 1,003 routine isolates, 73 anaerobes, and 53 bacterial enteropathogens, were tested on the Microflex LT and Axima Assurance devices. The spectra were analyzed using three databases: Biotyper (Bruker Daltonics), Saramis, and Vitek MS (bioMérieux). Among the routine isolates requiring identification to the species level (n = 986), 92.7% and 93.2% were correctly identified by the Biotyper and Vitek MS databases, respectively. The Vitek MS database is more specific
SUBMITTER: Martiny D
PROVIDER: S-EPMC3318529 | biostudies-literature | 2012 Apr