Stabilizing nanometer scale tip-to-substrate gaps in scanning electrochemical microscopy using an isothermal chamber for thermal drift suppression.


ABSTRACT: The control of a nanometer-wide gap between tip and substrate is critical for nanoscale applications of scanning electrochemical microscopy (SECM). Here, we demonstrate that the stability of the nanogap in ambient conditions is significantly compromised by the thermal expansion and contraction of components of an SECM stage upon a temperature change and can be dramatically improved by suppressing the thermal drift in a newly developed isothermal chamber. Air temperature in the chamber changes only at ~.2 mK/min to remarkably and reproducibly slow down the drift of tip-substrate distance to ~0.4 nm/min in contrast to 5-150 nm/min without the chamber. Eventually, the stability of the nanogap in the chamber is limited by its fluctuation with a standard deviation of ±0.9 nm, which is mainly as

SUBMITTER: Kim J 

PROVIDER: S-EPMC3328667 | biostudies-literature | 2012 Apr

REPOSITORIES: biostudies-literature