Unknown

Dataset Information

Perfusion cardiovascular magnetic resonance: Comparison of an advanced, high-resolution and a standard sequence.


ABSTRACT:

Background

Technical advances in perfusion cardiovascular magnetic resonance (CMR), particularly accelerated data acquisition methods, allow myocardial perfusion imaging with unprecedented spatial resolution. However, it is not clear how implementation of these recent advances affects perfusion image quality, signal and contrast to noise ratios (SNR & CNR) and the occurrence of important artefacts in routine clinical imaging. The objective of this study was therefore to compare a standard and an advanced, high-resolution perfusion sequence.

Methods

A standard ultrafast gradient echo perfusion sequence (st-GrE) was compared with an advanced kt-accelerated steady state free precession sequence (ktBLAST-SSFP) at 1.5 T in healthy volunteers (n = 16) and in patients (n = 32) with

SUBMITTER: Morton G 

PROVIDER: S-EPMC3403962 | biostudies-literature | 2012 Jun

REPOSITORIES: biostudies-literature