ABSTRACT: Background
The accuracy of continuous glucose monitoring (CGM) in non-critically ill hospitalized patients with heart failure or severe hyperglycemia (SH) is unknown.Methods
Hospitalized patients with congestive heart failure (CHF) exacerbation (receiving IV or subcutaneous insulin) or SH requiring insulin infusion were compared to outpatients referred for retrospective CGM.Results
Forty-three patients with CHF, 15 patients with SH, and 88 outpatients yielded 470, 164, and 2150 meter-sensor pairs, respectively. Admission glucose differed (188 versus 509 mg/dl in CHF and SH, p < .001) but not the first sensor glucose (p = .35). In continuous glucose error grid analysis, 67-78% of pairs during hypoglycemia were in zones A+B (p = .63), compared with 98-100% in euglycem