Unknown

Dataset Information

A comprehensive assessment of the precision and agreement of anterior corneal power measurements obtained using 8 different devices.


ABSTRACT:

Purpose

To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices.

Methods

Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Meditec, Germany), E300 (Medmont International, Australia), Allegro Topolyzer (Wavelight AG, Germany), Vista (EyeSys, TX), Pentacam (Oculus, Germany) and Sirius (CSO, Italy). Measurements were repeated in the second session (1 to 2 weeks later). Repeatability and reproducibility of corneal power measurements were assessed based on the intrasession and intersession within-subject standard de

SUBMITTER: Wang Q 

PROVIDER: S-EPMC3458095 | biostudies-literature | 2012

REPOSITORIES: biostudies-literature