Influence of sinogram affirmed iterative reconstruction of CT data on image noise characteristics and low-contrast detectability: an objective approach.


ABSTRACT:

Objectives

To utilize a novel objective approach combining a software phantom and an image quality metric to systematically evaluate the influence of sinogram affirmed iterative reconstruction (SAFIRE) of multidetector computed tomography (MDCT) data on image noise characteristics and low-contrast detectability (LCD).

Materials and methods

A low-contrast and a high-contrast phantom were examined on a 128-slice scanner at different dose levels. The datasets were reconstructed using filtered back projection (FBP) and SAFIRE and virtual low-contrast lesions (-20HU) were inserted. LCD was evaluated using the multiscale structural similarity index (MS-SIM*). Image noise texture and spatial resolution were objectively evaluated.

Results

The use of SAFIRE led to an improveme