Accuracy and reliability assessment of CT and MR perfusion analysis software using a digital phantom.


ABSTRACT:

Purpose

To design a digital phantom data set for computed tomography (CT) perfusion and perfusion-weighted imaging on the basis of the widely accepted tracer kinetic theory in which the true values of cerebral blood flow (CBF), cerebral blood volume (CBV), mean transit time (MTT), and tracer arrival delay are known and to evaluate the accuracy and reliability of postprocessing programs using this digital phantom.

Materials and methods

A phantom data set was created by generating concentration-time curves reflecting true values for CBF (2.5-87.5 mL/100 g per minute), CBV (1.0-5.0 mL/100 g), MTT (3.4-24 seconds), and tracer delays (0-3.0 seconds). These curves were embedded in human brain images. The data were analyzed by using 13 algorithms each for CT and magnetic resonance