Ontology highlight
ABSTRACT: Objective
To determine whether elevated blood cobalt (Co) concentrations are associated with early failure of metal-on-metal (MoM) hip resurfacings secondary to adverse reaction to metal debris (ARMD).Design
Cohort study.Setting
Single centre orthopaedic unit.Participants
Following the identification of complications potentially related to metal wear debris, a blood metal ion screening programme was instigated at our unit in 2007 for all patients with Articular Surface Replacement (ASR) and Birmingham MoM hip resurfacings. Patients were followed annually unless symptoms presented earlier. Symptomatic patients were investigated with ultrasound scan and joint aspiration. The clinical course of all 278 patients with 'no pain' or 'slight/occasional' pain and a H
SUBMITTER: Langton DJ
PROVIDER: S-EPMC3612810 | biostudies-literature | 2013 Mar