Unknown

Dataset Information

Same-day discovery of implantable cardioverter defibrillator dysfunction in the TRUST remote monitoring trial: influence of contrasting messaging systems.


ABSTRACT:

Aims

Assess whether automatic remote home monitoring (HM) permits same-day evaluation of implantable cardioverter defibrillator (ICD) system dysfunction.

Methods and results

Compromised ICD system integrity (generator/lead) demands prompt evaluation. Home monitoring promises earlier discovery but may be limited by technological differences and follow-up mechanism. We tested whether HM enabled event review within 24 h, and contrasted differing messaging mechanisms. Nine hundred and eight patients in the TRUST prospective multicentre trial were followed by HM for 15 months. ICD system problems automatically triggered notifications: repeatedly ('redundant') for impedance deviations and elective replacement indication (ERI), but only a single transmission for '30 J ineffective'.

SUBMITTER: Varma N 

PROVIDER: S-EPMC3636999 | biostudies-literature | 2013 May

REPOSITORIES: biostudies-literature