Limiting factors in atomic resolution cryo electron microscopy: no simple tricks.
Ontology highlight
ABSTRACT: To bring cryo electron microscopy (cryoEM) of large biological complexes to atomic resolution, several factors--in both cryoEM image acquisition and 3D reconstruction--that may be neglected at low resolution become significantly limiting. Here we present thorough analyses of four limiting factors: (a) electron-beam tilt, (b) inaccurate determination of defocus values, (c) focus gradient through particles, and (d) particularly for large particles, dynamic (multiple) scattering of electrons. We also propose strategies to cope with these factors: (a) the divergence and direction tilt components of electron-beam tilt could be reduced by maintaining parallel illumination and by using a coma-free alignment procedure, respectively. Moreover, the effect of all beam tilt components, including spira