Performance and Cost evaluation of one commercial and six in-house conventional and real-time reverse transcription-pcr assays for detection of severe acute respiratory syndrome coronavirus.


ABSTRACT: We evaluated seven reverse transcription-PCR (RT-PCR) assays, including six in-house assays and one commercial assay for the detection of severe acute respiratory syndrome coronavirus (SARS-CoV) RNA in clinical specimens. RT-PCR assays targeted different genomic regions and included three conventional assays (one nested and two non-nested) run on a conventional heat block and four real-time assays performed in a LightCycler (LC; Roche Diagnostics). All in-house assays were optimized for assay parameters, including MgCl2, primer, and probe concentrations. The commercial assay was the RealArt HPA CoV RT-PCR assay (Artus), which was run in the LC. Testing serial dilutions of cultured SARS-CoV showed that the analytical sensitivity of the assays ranged from 10(-8) to 10(-6), corresponding to 1

SUBMITTER: Mahony JB 

PROVIDER: S-EPMC387602 | biostudies-literature | 2004 Apr