Unknown

Dataset Information

SYBR green-based real-time quantitative PCR assay for detection of West Nile Virus circumvents false-negative results due to strain variability.


ABSTRACT: Real-time quantitative PCR is used routinely for the high-throughput diagnosis of viral pathogens, such as West Nile virus (WNV). Rapidly evolving RNA viruses present a challenge for diagnosis because they accumulate mutations that may render them undetectable. To explore the effect of sequence variations on assay performance, we generated every possible single point mutation within the target region of the widely used TaqMan assay for WNV and found that the TaqMan assay failed to detect 47% of possible single nucleotide variations in the probe-binding site and was unable to detect any targets with more than two mutations. In response, we developed and validated a less expensive assay with the intercalating dye SYBR green. The SYBR green-based assay was as sensitive as the TaqMan assay for

SUBMITTER: Papin JF 

PROVIDER: S-EPMC387603 | biostudies-literature | 2004 Apr

REPOSITORIES: biostudies-literature