Noninvasive skin fluorescence spectroscopy is comparable to hemoglobin A1c and fasting plasma glucose for detection of abnormal glucose tolerance.


ABSTRACT:

Aim

We compare performance of noninvasive skin fluorescence spectroscopy (SFS), fasting plasma glucose (FPG), and hemoglobin A1c (A1C) for detection of abnormal glucose tolerance (AGT).

Methods

The NSEEDS trial evaluated SFS, FPG, and A1C in an at-risk population of 479 previously undiagnosed subjects from nine US centers, each of whom received a 75 g, 2 h oral glucose tolerance test (OGTT). Skin fluorescence spectra were collected and analyzed with SCOUT DS® devices. Disease truth was AGT, defined as OGTT ≥140 mg/dl. Abnormal glucose tolerance sensitivity, false positive rate (FPR), and receiver operating characteristic (ROC) curves were computed for each measurement technique. Skin fluorescence spectroscopy reproducibility was also assessed.

Results

The AGT sensitiv

SUBMITTER: Olson BP 

PROVIDER: S-EPMC3879764 | biostudies-literature | 2013 Jul

REPOSITORIES: biostudies-literature