Performance comparisons of nano-LC systems, electrospray sources and LC-MS-MS platforms.
Ontology highlight
ABSTRACT: Selecting a suitable nano-liquid chromatography system (LC), ionization source and mass spectrometer for LC-tandem mass spectrometry (MS-MS) studies is complicated by numerous competing technologies. This study compares four popular nano-LC systems, four ionization sources and three MS facilities that use completely different LC-MS-MS systems. Statistically significant differences in LC performance were identified with similarly performing Proxeon, Waters and Eksigent nanoLC-Ultra systems [retention time routinely at 0.7-0.9% relative standard deviation (RSD)], and all outperformed the Eksigent nanoLC-2D (RSD ∼2%). In addition, compatibility issues were identified between the Bruker HCT ion trap mass spectrometer and both the Eksigent nanoLC-2D and the Bruker nanoelectrospray source. The e
SUBMITTER: Liu Q
PROVIDER: S-EPMC3886563 | biostudies-literature | 2014 Feb