Cardiac resynchronization therapy and AV optimization increase myocardial oxygen consumption, but increase cardiac function more than proportionally.


ABSTRACT:

Background

The mechanoenergetic effects of atrioventricular delay optimization during biventricular pacing ("cardiac resynchronization therapy", CRT) are unknown.

Methods

Eleven patients with heart failure and left bundle branch block (LBBB) underwent invasive measurements of left ventricular (LV) developed pressure, aortic flow velocity-time-integral (VTI) and myocardial oxygen consumption (MVO2) at 4 pacing states: biventricular pacing (with VV 0 ms) at AVD 40 ms (AV-40), AVD 120 ms (AV-120, a common nominal AV delay), at their pre-identified individualised haemodynamic optimum (AV-Opt); and intrinsic conduction (LBBB).

Results

AV-120, relative to LBBB, increased LV developed pressure by a mean of 11(SEM 2)%, p=0.001, and aortic VTI by 11(SEM 3)%, p=0.002, but also