Risk Minimization Measures for Blood Screening HIV-1 Nucleic Acid Amplification Technique Assays in Germany.


ABSTRACT:

Background

Several publications describe HIV-1 RNA false-negative results or viral load underquantitation associated with Communauté Européenne(CE)-marked qualitative or quantitative nucleic acid amplification technique (NAT) assays. 6 cases occurred during blood screening in Germany, with 2 of them causing HIV-1 transmissions to recipients of blood components. The implicated NAT assays were mono-target assays amplifying in different viral genome regions (gag or long terminal repeat).

Methods

Specimens characterized by HIV-1 NAT underquantitation or false-negative NAT results were comparatively investigated in CE-marked HIV-1 NAT systems of different design to identify potential reasons. The target regions of the viral nucleic acids were sequenced and these sequences compare