Developing a class solution for Prostate Stereotactic Ablative Body Radiotherapy (SABR) using Volumetric Modulated Arc Therapy (VMAT).


ABSTRACT:

Background and purpose

To develop a class solution for prostate Stereotactic Ablative Radiotherapy (SABR) using Volumetric Modulated Arc Therapy (VMAT).

Materials and methods

Seven datasets were used to compare plans using one 360° arc (1FA), one 210° arc (1PA), two full arcs and two partial arcs. Subsequently using 1PA, fifteen datasets were compared using (i) 6mm CTV-PTV margins, (ii) 8mm CTV-PTV margins and (iii) including the proximal SV within the CTV. Monaco 3.2 (Elekta) was used for planning with the Agility MLC system (Elekta).

Results

Highly conformal plans were produced using all four arc arrangements. Compared to 1FA, 1PA resulted in significantly reduced rectal doses, and monitor units and estimated delivery times were reduced in six of seven cases. Using