Unknown

Dataset Information

Dual-beam histotripsy: a low-frequency pump enabling a high-frequency probe for precise lesion formation.


ABSTRACT: Histotripsy produces tissue fractionation through dense energetic bubble clouds generated by short, high-pressure, ultrasound pulses. When using pulses shorter than 2 cycles, the generation of these energetic bubble clouds only depends on where the peak negative pressure (P-) exceeds the intrinsic threshold of the medium (26 to 30 MPa in soft tissue with high water content). This paper investigates a strategic method for precise lesion generation in which a low-frequency pump pulse is applied to enable a sub-threshold high-frequency probe pulse to exceed the intrinsic threshold. This pump-probe method of controlling a supra-threshold volume can be called dual-beam histotripsy. A 20-element dual-frequency (500-kHz and 3-MHz elements confocally aligned) array transducer was used to generate

SUBMITTER: Lin KW 

PROVIDER: S-EPMC3971546 | biostudies-literature | 2014 Feb

REPOSITORIES: biostudies-literature