New shielding configurations for a simultaneous PET/MRI scanner at 7T.
Ontology highlight
ABSTRACT: Understanding sources of electromagnetic interference are important in designing any electronic system. This is especially true when combining positron emission tomography (PET) and magnetic resonance imaging (MRI) in a multimodality system as coupling between the subsystems can degrade the performance of either modality. For this reason, eliminating radio frequency (RF) interference and gradient-induced eddy currents have been major challenges in building simultaneous hybrid PET/MRI systems. MRI requires negligible RF interference at the Larmor resonance frequency, while RF interference at almost any frequency may corrupt PET data. Moreover, any scheme that minimizes these interactions would, ideally, not compromise the performance of either subsystem. This paper lays out a plan to resolv
SUBMITTER: Peng BJ
PROVIDER: S-EPMC3976213 | biostudies-literature | 2014 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA