Unknown

Dataset Information

A 1.5T MRI-conditional 12-lead electrocardiogram for MRI and intra-MR intervention.


ABSTRACT:

Purpose

High-fidelity 12-lead electrocardiogram (ECG) is important for physiological monitoring of patients during MR-guided intervention and cardiac MRI. Issues in obtaining noncorrupted ECGs inside MRI include a superimposed magneto-hydro-dynamic voltage, gradient switching-induced voltages, and radiofrequency heating. These problems increase with magnetic field. The aim of this study is to develop and clinically validate a 1.5T MRI-conditional 12-lead ECG system.

Methods

The system was constructed with transmission lines to reduce radiofrequency induction and switching circuits to remove induced voltages. Adaptive filters, trained by 12-lead measurements outside MRI and in two orientations inside MRI, were used to remove the magneto-hydro-dynamic voltage. The system was t

SUBMITTER: Tse ZT 

PROVIDER: S-EPMC3976440 | biostudies-literature | 2014 Mar

REPOSITORIES: biostudies-literature