Evaluation of the Bruker MALDI Biotyper for identification of Gram-positive rods: development of a diagnostic algorithm for the clinical laboratory.
Ontology highlight
ABSTRACT: Reported matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) identification rates of Gram-positive rods (GPR) are low compared to identification rates of Gram-positive cocci. In this study, three sample preparation methods were compared for MALDI-TOF MS identification of 190 well-characterized GPR strains: direct transfer, direct transfer-formic acid preparation, and ethanol-formic acid extraction. Using the interpretation criteria recommended by the manufacturer, identification rates were significantly higher for direct transfer-formic acid preparation and ethanol-formic acid extraction than for direct transfer. Reducing the species cutoff from 2.0 to 1.7 significantly increased species identification rates. In a subsequent prospective study, 215 cl
SUBMITTER: Schulthess B
PROVIDER: S-EPMC3993486 | biostudies-literature | 2014 Apr
REPOSITORIES: biostudies-literature
ACCESS DATA