Quantification of proton dose calculation accuracy in the lung.


ABSTRACT:

Purpose

To quantify the accuracy of a clinical proton treatment planning system (TPS) as well as Monte Carlo (MC)-based dose calculation through measurements and to assess the clinical impact in a cohort of patients with tumors located in the lung.

Methods and materials

A lung phantom and ion chamber array were used to measure the dose to a plane through a tumor embedded in the lung, and to determine the distal fall-off of the proton beam. Results were compared with TPS and MC calculations. Dose distributions in 19 patients (54 fields total) were simulated using MC and compared to the TPS algorithm.

Results

MC increased dose calculation accuracy in lung tissue compared with the TPS and reproduced dose measurements in the target to within ±2%. The average difference be