Rapid flow-based peptide synthesis.
Ontology highlight
ABSTRACT: A flow-based solid-phase peptide synthesis methodology that enables the incorporation of an amino acid residue every 1.8 min under automatic control or every 3 min under manual control is described. This is accomplished by passing a stream of reagent through a heat exchanger into a low volume, low backpressure reaction vessel, and through a UV detector. These features enable continuous delivery of heated solvents and reagents to the solid support at high flow rate, thereby maintaining maximal concentration of reagents in the reaction vessel, quickly exchanging reagents, and eliminating the need to rapidly heat reagents after they have been added to the vessel. The UV detector enables continuous monitoring of the process. To demonstrate the broad applicability and reliability of this method
SUBMITTER: Simon MD
PROVIDER: S-EPMC4045704 | biostudies-literature | 2014 Mar
REPOSITORIES: biostudies-literature
ACCESS DATA