Effect of inhaler design variables on paediatric use of dry powder inhalers.


ABSTRACT: Age appropriateness is a major concern of pulmonary delivery devices, in particular of dry powder inhalers (DPIs), since their performance strongly depends on the inspiratory flow manoeuvre of the patient. Previous research on the use of DPIs by children focused mostly on specific DPIs or single inspiratory parameters. In this study, we investigated the requirements for a paediatric DPI more broadly using an instrumented test inhaler. Our primary aim was to assess the impact of airflow resistance on children's inspiratory flow profiles. Additionally, we investigated children's preferences for airflow resistance and mouthpiece design and how these relate to what may be most suitable for them. We tested 98 children (aged 4.7-12.6 years), of whom 91 were able to perform one or more correct in