Improved measurement of the rotor temperature in analytical ultracentrifugation.
Ontology highlight
ABSTRACT: Sedimentation velocity is a classical method for measuring the hydrodynamic, translational friction coefficient of biological macromolecules. In a recent study comparing various analytical ultracentrifuges, we showed that external calibration of the scan time, radial magnification, and temperature is critically important for accurate measurements (Anal. Biochem. 440 (2013) 81-95). To achieve accurate temperature calibration, we introduced the use of an autonomous miniature temperature logging integrated circuit (Maxim Thermochron iButton) that can be inserted into an ultracentrifugation cell assembly and spun at low rotor speeds. In the current work, we developed an improved holder for the temperature sensor located in the rotor handle. This has the advantage of not reducing the rotor capa
SUBMITTER: Zhao H
PROVIDER: S-EPMC4051143 | biostudies-literature | 2014 Apr
REPOSITORIES: biostudies-literature
ACCESS DATA