Commutability of possible external quality assessment materials for cardiac troponin measurement.


ABSTRACT:

Background

The measurement of cardiac troponin is crucial in the diagnosis of myocardial infarction. The performance of troponin measurement is most conveniently monitored by external quality assessment (EQA) programs. The commutability of EQA samples is often unknown and the effectiveness of EQA programs is limited.

Methods

Commutability of possible EQA materials was evaluated. Commercial control materials used in an EQA program, human serum pools prepared from patient samples, purified analyte preparations, swine sera from model animals and a set of patient samples were measured for cTnI with 4 assays including Abbott Architect, Beckman Access, Ortho Vitros and Siemens Centaur. The measurement results were logarithm-transformed, and the transformed data for patient samples