Detection of significant bacteriuria by use of the iQ200 automated urine microscope.


ABSTRACT: In the microbiology laboratory, there is an augmented need for rapid screening methods for the detection of bacteria in urine samples, since about two-thirds of these samples will not yield any bacteria or will yield insignificant growth when cultured. Thus, a reliable screening method can free up laboratory resources and can speed up the reporting of a negative urine result. In this study, we have evaluated the detection of leukocytes, bacteria, and a new sediment indicator, the "all small particles" (ASP), by an automated instrument, the iQ200 urine analyzer, to detect negative urine samples that can be excluded from culture. A coupled automated strip reader (iChem Velocity), enabling the detection of nitrite and leukocyte esterase, was tested in parallel. In total, 963 urine samples wer