Automatic scoring of sleep stages and cortical arousals using two electrodes on the forehead: validation in healthy adults.
Ontology highlight
ABSTRACT: Accuracy and limitations of automatic scoring of sleep stages and electroencephalogram arousals from a single derivation (Fp1 -Fp2 ) were studied in 29 healthy adults using a portable wireless polysomnographic recorder. All recordings were scored five times: twice by a referent scorer who viewed the standard polysomnographic montage and observed the American Academy of Sleep Medicine rules (referent scoring and blind rescoring); and once by the same scorer who viewed only the Fp1 -Fp2 signal (alternative scoring), by another expert from the same institution, and by the algorithm. Automatic, alternative and independent expert scoring were compared with the referent scoring on an epoch-by-epoch basis. The algorithm's agreement with the reference (81.0%, Cohen's κ = 0.75) was comparable to th