ABSTRACT: Background
The use of a video-assisted laryngoscope (VL) has been shown to reduce the time to achieve intubation with a double-lumen endobronchial tube (DLT). As the blade of the VL is curved differently to a standard laryngoscope, the DLT must be angled into a hockey stick shape to fit properly. We conducted a study to establish which direction of angulation was best to facilitate correct positioning of the DLT when using a VL.Methods
We enrolled patients scheduled for thoracic surgery who required intubation with a DLT. They were prospectively randomized into one of two groups: those intubated with a DLT angled to conceal the tracheal orifice (the tracheal orifice-covered, TOC) group or the tracheal orifice-exposed (TOE) group. The composite primary outcome measures were