ABSTRACT: Background
Continuous glucose monitoring (CGM) is a tool widely used in the treatment of patients with type 1 diabetes. The purpose of the current study was to evaluate whether accuracy and patient treatment satisfaction differ between the Enlite™ (Medtronic MiniMed, Inc., Northridge, CA) and Dexcom(®) (San Diego, CA) G4 PLATINUM CGM sensors.Subjects and methods
Thirty-eight ambulatory patients with type 1 diabetes used the Dexcom G4 and Enlite sensors simultaneously for a minimum of 4 and maximum of 6 days. Patients measured capillary glucose levels with a HemoCue(®) (Ängelholm, Sweden) system six to 10 times a day. In addition, two inpatient studies were performed between Days 1-3 and 4-6.Results
The mean absolute relative difference (MARD) in blood glucose for th