Quantitative and qualitative assessment of Yttrium-90 PET/CT imaging.


ABSTRACT: Yttrium-90 is known to have a low positron emission decay of 32 ppm that may allow for personalized dosimetry of liver cancer therapy with (90)Y labeled microspheres. The aim of this work was to image and quantify (90)Y so that accurate predictions of the absorbed dose can be made. The measurements were performed within the QUEST study (University of Sydney, and Sirtex Medical, Australia). A NEMA IEC body phantom containing 6 fillable spheres (10-37 mm ∅) was used to measure the 90Y distribution with a Biograph mCT PET/CT (Siemens, Erlangen, Germany) with time-of-flight (TOF) acquisition. A sphere to background ratio of 8:1, with a total (90)Y activity of 3 GBq was used. Measurements were performed for one week (0, 3, 5 and 7 d). he acquisition protocol consisted of 30 min-2 bed positions